4 foil clamp type 2 (optional), Operation – Leica Biosystems RM2125 RTS User Manual
Page 30

30
Instructions for Use, V2.1 – 10/2012
• To open the jaws (56) and (57), slightly loos-
en the three screws (55) with an Allen key
No. 4 (
49
).
• Insert the foil ribbon (54) from behind to po-
sition it between the movable jaw (57) and
the fixed jaw (56).
• To clamp the foil, first tighten the screw in
the middle and then the other two screws
(55) with the Allen key (49).
• Place the foil clamp (59) in the standard
specimen clamp so that the beveled surface
(58) on the back of the foil clamp points to
the right or left.
• Turn the knurled screw (40) clockwise until
the foil clamp is clamped securely.
5.9.4 Foil clamp Type 2
(optional)
The foil clamp type 2 is appropriate for large foil ribbons.
It is mounted in the standard specimen clamp.
Fig. 23
49
54
57
56
40
55
58
59
5. Operation
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)