Startup, 1 installation site requirements, 2 standard delivery – Leica Biosystems RM2125 RTS User Manual
Page 14: Installation site requirements, Standard delivery

14
Instructions for Use, V2.1 – 10/2012
4. Startup
4.2 Standard delivery
4.1 Installation site requirements
- The instrument requires an installation area of approx. 438 x 472 mm.
- Room temperature consistently between +10 °C and +35 °C
- Relative humidity 80 % maximum - non-condensing
- Ambient pressure from 740 hPa to 1100 hPa
- Elevation: up to a maximum of 2000 m above sea level
- The instrument is designed for indoor use only.
- Obstruction-free access to the handwheel.
- For smooth function, also make sure that no other instruments that cause vibrations are
standing in the immediate vicinity of the instrument.
- The counter top must be largely vibration-free and have sufficient load capacity and rigid-
ity for the weight of the instrument.
- Avoid vibrations, direct sunlight, and large temperature fluctuations.
- The chemicals to be used are easily inflammable and hazardous to health. Therefore the
installation location must be well ventilated, and must contain no sources of ignition of
any kind.
1 Leica RM2125 RTS basic instrument ............................................................................14 0457 46960
The basic instrument includes the following:
1 specimen holder fixture, directional .............................................................................14 0457 46961
with quick change system for specimen clamps (installed on the instrument)
1 tool set - consisting of:
1 Allen key with handle, No. 3 .....................................................................................14 0194 04764
1 Allen key with handle, No. 4 .....................................................................................14 0194 04782
1 Allen key, No. 8 ...........................................................................................................14 0222 04143
1 bottle (50 ml) of oil for drives, type 405 ...................................................................14 0336 06086
1 dust cover (104 R) .......................................................................................................14 0212 04091
1 set of Instructions for Use RM2125RTS (+ language CD) ..........................................14 0457 80002
You will find this, as well as other accessories you may have ordered, packed in the top of the
carton (
see Fig. 7
).
Compare the delivered components with the parts list and your order.
Should there be any discrepancy, please contact the Leica selling unit handling your