3 technical data, Technical data, Instrument components and specifications – Leica Biosystems RM2125 RTS User Manual
Page 13

13
Leica RM2125RTS
3.3 Technical Data
Installation site requirements
Operating temperature range: 10 °C- 35 °C
Temperature range during storage: 5 °C - 55 °C
Relative humidity:
max. 80 %, non-condensing
Storage
humidity:
<
80
%
General
Section thickness range: 0.5 - 60µm
Section thickness settings: from 0 - 2 µm in 0.5 µm increments
from 2 - 10 µm in 1 µm increments
from 10 - 20 µm in 2 µm increments
from 20 - 60 µm in 5 µm increments
Object
feed:
25
mm
Vertical
stroke:
59
mm
Specimen
retraction: ON/OFF
Dimensions and weight
Width
438
mm
Depth
472
mm
Height
265
mm
Working height (knife blade) 105 mm
Weight (without accessories) 29 kg
Optional equipment and optional accessories
Specimen orientation (option)
Horizontal:
8°
Vertical:
8°
Angle of rotation:
± 90°
Trimming stages:
10 µm 50 µm
Repositioning of knife holder base
with lateral movement
North-south:
±
25
mm
East-west movement: ± 20 mm
without lateral movement
North-south:
±
25
mm
3. Instrument Components and Specifications