6 attaching the knife holder base, non-orientable, 7 inserting the knife holder, non-orientable, Attaching the knife holder base, non-orientable – Leica Biosystems RM2125 RTS User Manual
Page 18: Inserting the knife holder, non-orientable, Startup

18
Instructions for Use, V2.1 – 10/2012
4. Startup
• Release the clamping lever (20) by rotating it
counterclockwise.
(Position
= released)
• Insert the universal knife holder base (21)
using the notch (22) on the bottom into the
T-piece (19) of the microtome base plate
(23).
• To secure the knife holder base, turn the
clamping lever (20) clockwise.
(Position
= locked)
• Loosen the Allen screw (24) by turning it
counterclockwise.
• Slide the notch of the knife holder (25) onto
the T-piece (26) of the knife holder base (21).
• To clamp, turn the Allen screw (24) clock-
wise.
Fig. 10
4.6 Attaching the knife holder base, non-orientable
Fig. 11
4.7 Inserting the knife holder, non-orientable
19
20
21
26
25
22
23
24
21
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)