8 Configuration

8.5 Decimal point (decimal place)

DP

Select the position of the decimal point.

Select between:

0.1 = one decimal place and
1 = no decimal place

factory-set:

0.1

h Press the UP ( ) or DOWN ( ) key to change between the formats.
h Press the SET key to activate the selected format.
h Press the ESC key twice to return to normal display.

8.6 Temperature correction

PV Bias

Using this parameter, the temperature can be corrected by a constant value,
e.g. with known measurement errors.

Select between:

-99.9 to +99.9 °C (°F)

factory-set:

0.0

h Select the digit to be altered by pressing the LEFT ( ) or RIGHT ( ) key

(digit is shown blinking).

h Press the UP ( ) or DOWN ( ) key to alter the value.
h Press the SET key to activate the selected correction.
h Press the ESC key twice to return to normal display.

Example

“PV Bias” is set to 1.0°C. The temperature is 99.0°C. The sum of both values
is displayed and stored (100.0°C).

blinking

blinking

Measured
temperature
99.0°C

PV Bias
1.0°C