beautypg.com

2 automatic logging, 1 activating the interval, Automatic logging – JUMO 702540 TDA-300 / TDA-3000 - Handheld thermometer Operating Manual User Manual

Page 33: 5 data logger

background image

33

5 Data logger

5.2 Automatic logging

The measured data are recorded in the memory at the preset interval. For
automatic data logging, you must first set the interval.

The following can be logged:
- the measured temperature,
- the date (without “year”),
- the TAG name and
- the user name.

5.2.1 Activating the interval

h In normal display, press the MENU/POWER key

three times to access the function menu
“Logging (Fn3)”.

h Press the SET key.

You will now see the interval setting screen.

Interval:

setting range: OFF (manual logging) or

1 to 3600 seconds (automatic logging)

factory-set: OFF

h Set the interval with the UP key

(

)

and the shift

keys

(

,

)

.

Example: interval 15 seconds

h Press the SET key to save the new value.

The display returns to the data logging screen
(Fn3). When you have set the interval (to a value

other than “OFF”), the REC symbol (

) blinks

to indicate “ready to measure”; automatic
measurement, however, is not yet active.

H

To return to normal display after setting, press the ESC key.

blinking

blinking