12.3. List of commands
out commands: Setting parameters or temperature values.
Command
Parameter
Response of circulator
version
None
Number of software version (V X.xx)
status
none
Status message, error message (see page 62 )
out_mode_01
0
Use working temperature >t 1<
out_mode_01
1
Use working temperature >t 2<
out_mode_01
2
Use working temperature >t 3<
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_sp_00
xxx.xx
Set working temperature. „t 1“
out_sp_01
xxx.xx
Set working temperature. „t 2“
out_sp_02
xxx.xx
Set working temperature. „t 3“
out_sp_03
xxx.xx
Set high temperature warning limit
„t High“
out_sp_04
xxx.xx
Set low temperature warning limit
„t Low“
out_sp_07
x
Set the pump pressure stage. (1 ... 4)
out_par_06
xxx
Xp control parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.
in commands: Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04
none
Setpoint temperature („SafeTemp“) of the excess
temperature protection
in_sp_00
none
Working temperature „t 1“
in_sp_01
none
Working temperature „t 2“
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators