12.3. List of commands

out commands: Setting parameters or temperature values.

Command

Parameter

Response of circulator

version

None

Number of software version (V X.xx)

status

none

Status message, error message (see page 62 )

out_mode_01

0

Use working temperature >t 1<

out_mode_01

1

Use working temperature >t 2<

out_mode_01

2

Use working temperature >t 3<

out_mode_05

0

Stop the unit = R –OFF-.

out_mode_05

1

Start the unit.

out_sp_00

xxx.xx

Set working temperature. „t 1“

out_sp_01

xxx.xx

Set working temperature. „t 2“

out_sp_02

xxx.xx

Set working temperature. „t 3“

out_sp_03

xxx.xx

Set high temperature warning limit

„t High“

out_sp_04

xxx.xx

Set low temperature warning limit

„t Low“

out_sp_07

x

Set the pump pressure stage. (1 ... 4)

out_par_06

xxx

Xp control parameter of the internal controller.

out_par_07

xxx

Tn control parameter of the internal controller.

out_par_08

xxx

Tv control parameter of the internal controller.



in commands: Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of circulator

in_pv_00

none

Actual bath temperature.

in_pv_01

none

Heating power being used (%).

in_pv_03

none

Temperature value registered by the safety sensor.

in_pv_04

none

Setpoint temperature („SafeTemp“) of the excess
temperature protection

in_sp_00

none

Working temperature „t 1“

in_sp_01

none

Working temperature „t 2“