beautypg.com

JULABO FPW50-MA Refrigerated and Heating Circulators User Manual

Page 21

background image

MA

21

MA

Temperature selection

digital

via keypad

indication on

MULTI-DISPLAY(LED)

remote control via personal computer

indication on monitor

Temperature indication

MULTI-DISPLAY (LED)

Resolution

(-9.99 …. +99.99 = 0.01)

°C

0.01 / 0.1

Absolute Temperature Calibration

°C

±3

Temperature control

PID

Heater wattage

(at 230 V)

kW

2,0

Heater wattage

(at 115V)

kW

1,0

Electronically adj. pump capacity

stages

1 ... 4

Flow rate

max.at 0 bar

l/min

11 ... 16

Pressure max.

max. at 0 l

bar

0.23 ... 0.45

Electrical connections:

External alarm device

Vdc/mA

24-0 / max. 25

Computer interface

RS232

Ambient temperature

°C

5 ... 40


Safety installations according to IEC 61010-2-010:

Excess temperature protection

adjustable from 0 °C ... 230 °C

Low liquid level protection

float switch

Classification according to DIN 12876-1

class III

Supplementary safety installations

Early warning system for low level

float switch

High temperature warning function

optical + audible (in intervals)

Low temperature warning function

optical + audible (in intervals)

Supervision of working sensor

plausibility control

Reciprocal sensor monitoring between
working and safety sensors

difference >35 K

Alarm message

optical + audible (permanent)

Warning message

optical + audible (in intervals)