beautypg.com

JULABO FPW50-MA Refrigerated and Heating Circulators User Manual

Page 19

background image

MA

19

F33-MA

F34-MA

Working temperature range

°C

-30 ... 200

-30 ... 150

Temperature stability

°C

±0,02

±0,02

Cooling capacity

Medium ethanol

°C
kW

+20 0 -20 -30
0.5 0.32 0.12 0.03

+20 0 -20 -30

0.45 0.32 0.14 0.03

Refrigerant

R134a

R134a

Overall dimensions

(WxDxH)

cm

36x46x69

38x58x62

Bath opening

(WxL)

cm

23x14

24x30

Bath depth

cm

20

15

Filling volume

from ... to

liters

12 ... 16

14 ... 20

Weight

kg

44

42

Mains power connection

230 V/50 Hz

V/ Hz 207-253 / 50

207-253 / 50

Current draw

(at 230 V)

A

12

12

Current draw

CH (at 230 V)

A

9 + 2

9 + 2

Mains power connection

230 V/60 Hz

V/ Hz 207-253 / 60

207-253 / 60

Current draw

(at 230 V)

A

12

12

Current draw

(at 208 V)

A

12

13

Mains power connection

115 V/60 Hz

V/ Hz 103-127 / 60

103-127 / 60

Current draw

(at 115 V)

A

15

14

Mains power connection 100 V/50-60 Hz

V/ Hz 90-110/50-60Hz

-----

Current draw

(at 100 V)

A

15

-----



FP35-MA

Working temperature range

°C

-35 ... 150

Temperature stability

°C

±0,02

Cooling capacity

Medium ethanol

°C
kW

+20 0 -20 -30
0.45 0.39 0.15 0.05

Refrigerant

R134a

Overall dimensions

(WxDxH)

cm

31x42x64

Bath opening

(WxL)

cm

18x12

Bath depth

cm

5

Filling volume

from ... to

liters

1,7 ... 2.5

Weight

kg

37

Mains power connection

230 V/50 Hz

V/ Hz 207-253 / 50

Current draw

(at 230 V)

A

12

Current draw

CH (at 230 V)

9 / 2

Mains power connection

115 V/60 Hz

V/ Hz 103-127 / 60

Current draw

(at 115 V)

A

14

Mains power connection 100 V/50-60 Hz

V/ Hz 90-110/50-60Hz

Current draw

(at 100 V)

A

14


All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C

Technical changes without prior notification reserved.