beautypg.com

JULABO FPW50-MA Refrigerated and Heating Circulators User Manual

Page 54

background image

Menu functions

54

The value >tt1< is only indicated

Setting >Ct 1< by using the keys.

• Integer digits flash, set by pressing

(79) +

• Decimal digits flash, set by pressing

(70) +

The first of 3 points is calibrated.

Return to 1. Set working temperature value: 120.00 °C

The value >tt2< is only indicated

Setting >Ct 2< by using the keys.

• Integer digits flash, set by pressing

(119) +

• Decimal digits flash, set by pressing

(5) +

The second of 3 points is calibrated.

Return to 1. Set working temperature value: 160.00 °C

The value >tt3< is only indicated

Setting >Ct 3< by using the keys.

• Integer digits flash, set by pressing

(159) +

• Decimal digits flash, set by pressing

(3) +

The 3-point calibration is completed

4. Return to standard operation

• Set >YES< after calibration.

(Standard operation)