Operator responsibility – Safety recommendations

20

FP40-MA

Working temperature range

°C

-40 ... 200

Temperature stability

°C

±0,02

Cooling capacity

Medium ethanol

°C
kW

+20 0 -20 -30 -40
0.68 0.5 0.32 0.17 0.04

Refrigerant

R404A

Overall dimensions

(WxDxH)

cm

37x46x69

Bath opening

(WxL)

cm

23x14

Bath depth

cm

20

Filling volume

from ... to

liters

9 ... 16

Weight

kg

48

Mains power connection

230 V/50 Hz

V/ Hz

207-253 / 50

Current draw

(at 230 V)

A

13

Current draw

CH (at 230 V)

A

9+3

Mains power connection

230 V/60 Hz

V/ Hz

207-253 / 60

Current draw

(at 230 V)

A

13







FP50-MA / FPW50-MA

Working temperature range

°C

-50 ... 200

Temperature stability

°C

±0,02

Cooling capacity

Medium ethanol

°C
kW

+20 0 -20 -30 -40
0.9 0.8 0.5 0.32 0.16

Refrigerant

R404A / R507

Overall dimensions

(WxDxH)

cm

42x49x70

Bath opening

(WxL)

cm

18x12

Bath depth

cm

15

Filling volume

from ... to

liters

5,5 ... 8

Weight

kg

55

Mains power connection

230 V/50 Hz

V/ Hz

207-253 / 50

Current draw

(at 230 V)

A

14

Current draw

CH (at 230 V)

A

9+4

Mains power connection

230 V/60 Hz

V/ Hz

207-253 / 60

Current draw

(at 230 V)

A

14


All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C

Technical changes without prior notification reserved.