Reset - factory setting, Menu control parameters – xp, tv, tn – JULABO FPW50-MA Refrigerated and Heating Circulators User Manual
Page 47
MA
47
Warning
:
For supervised or unsupervised operation with the “AUTOSTART“ function
avoid any hazardous situation to persons or property
Take care to fully observe the safety and warning functions of the circulator.
9.2.4. Reset - Factory setting
>YES< resets all values to factory setting.
A >rSt< RESET can be effected in the >
< mode only.
Switch off the circulator by pressing the key
and call up the
menu CONFIGURATION.
During the message – init - all parameters are reset to factory
settings
9.3.
MENU Control parameters – Xp, Tv, Tn
Menu level 1
In most cases the control parameters preset in the factory are adequate
for achieving an optimum temperature sequence.
The control parameters allow adjustment to special control processes..
Level 2
Parameter level
Press the
key if a parameter is to be
retained. Correction function for parameters or
values (prior to OK)
0.1 … 99.9
Proportional range >Xp<
•
The parameter flashes, switch by pressing
and
3 … 9999
Reset time >Tn< (Integral component)
•
The parameter flashes, switch by pressing
and
0 … 999
Lead time >Tv< (Differential component)
•
The parameter flashes, switch by pressing
and
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators