Technical specifications – JULABO FPW50-MA Refrigerated and Heating Circulators User Manual
Page 18

Operator responsibility – Safety recommendations
18
2.4.
Technical specifications
F12-MA
F25-MA
Working temperature range
°C
-20 ... 200
-28 ... 200
Temperature stability
°C
±0,02
±0,02
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.16 0.1 0.02
+20 0 -20
0.26 0.2 0.06
Refrigerant
R134a
R134a
Overall dimensions
(WxDxH)
cm
20x36x56
23x42x61
Bath opening
(WxL)
cm
13x15
12x14
Bath depth
cm
13
14
Filling volume
from ... to
liters
3 ... 4,5
3 ... 4,5
Weight
kg
23
31
Mains power connection
230 V/50 Hz
V/ Hz
207-253 / 50
207-253 / 50
Current draw
(at 230 V)
A
11
12
Current draw
CH (at 230 V)
A
9 + 1
9 + 2
Mains power connection
230 V/60 Hz
V/ Hz
207-253 / 60
207-253 / 60
Current draw
(at 230 V)
A
11
12
Current draw
(at 208 V)
A
11
12
Mains power connection
115 V/60 Hz
V/ Hz
103-127 / 60
103-127 / 60
Current draw
(at 115 V)
A
12
13
Mains power connection 100 V/50-60 Hz
V/ Hz
90-120 / 50-60
90-115 / 50-60
Current draw
(at 100 V)
A
15
13
F32-MA
Working temperature range
°C
-35 ... 200
Temperature stability
°C
±0,02
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.45 0.39 0.15
Refrigerant
R134a
Overall dimensions
(WxDxH)
cm
31x42x64
Bath opening
(WxL)
cm
18x12
Bath depth
cm
15
Filling volume
from ... to
liters
5,5 ... 8
Weight
kg
37
Mains power connection
230 V/50 Hz
V/ Hz
207-253 / 50
Current draw
(at 230 V)
A
12
Current draw
CH (at 230 V)
A
9 + 2
Mains power connection
230 V/60 Hz
V/ Hz
207-253 / 60
Current draw
(at 230V)
A
12
Current draw
(at 208V)
A
10
Mains power connection
115 V/60 Hz
V/ Hz
103-127 / 60
Current draw
(at 115 V)
A
14
Mains power connection 100 V/50-60 Hz
V/ Hz
90-115 / 50-60
Current draw
(at 100 V)
A
14
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators