JULABO FPW50-MA Refrigerated and Heating Circulators User Manual
Page 51

MA
51
>tYP< - ATC TYPE
•
The parameter flashes, switch by pressing
and
A >1-point<, >2-point< or >3-point<
calibration can be carried out.
The value > tt1< is only indicated
In addition the measured temperature value
>Ct 1< is saved during the next step.
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
If only a 1-point calibration is carried out, the following menu items
are not indicated anymore
The value > tt2< is only indicated
In addition the measured temperature value
>Ct 2< is saved during the next step.
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
If only a 2-point calibration is carried out, the following menu items
are not indicated anymore
The value > tt3< is only indicated
In addition the measured temperature value
>Ct 3< is saved during the next step.
•
Integer digits flash, set by pressing
+
•
Decimal digits flash, set by pressing
+
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators