Operator responsibility – Safety recommendations

22

Environmental conditions according to IEC 61 010-1:

Use only indoor.

Altitude up to 2000 m - normal zero.

Ambient temperature: +5 ... +40 °C

Air humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,

linear decrease down to 50 % relative humidity at a temperature of +40 °C

Max. mains fluctuations of ±10 % are permissible.

Protection class according to IEC 60 529

IP21

The unit corresponds to Class I

Overvoltage category

II

Pollution degree

2

Caution:

The unit is not for use in explosive environment


Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes)
Class A (industrial and commercial range).



2.5.

Cooling water connection

Cooling water pressure (IN/OUT)

max.

6 bar

Pressure difference (IN - OUT)

3.5 to 6 bar

Cooling water temperature

< 20 °C