JULABO FPW50-MA Refrigerated and Heating Circulators User Manual
Page 58

Troubleshooting guide / error messages
58
„Configuration Error“
The configuration of the circulator does not correspond with its current
application.
Press the
key for a non-recurring, automatic change of the
configuration.
In this case please call the JULABO Technical Service or an authorized
dealer.
Disturbances that are not indicated.
The electronic pump motor is overload-protected by an electronic current
limiter. If viscosity of the bath fluid is or becomes too high, the motor
stops running.
15
A
M
P
Mains circuit breakers (resettable) 15 A.
Cooling machine: Fuse T 10.0 A, dia.5 x 20 mm
The mains fuses (8b) on the rear of the unit may easily be exchanged
as shown on the left.
Warning:
Before exchanging the fuses, turn off the mains power switch and disconnect
the power plug from the mains socket!
Only use fine fuses with a nominal value as specified.
Example:
Manufacturer
Supplier
Type
Order No.
Wickmann
Wickmann
G- fuse insert
T10,0A 5x20 mm
No. 19195
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators