• Do not stay in the area below the unit.
• Make sure you read and understand all instructions and safety precautions
listed in this manual before installing or operating your unit.
• Set the excess temperature safety installation at least 25 °C below the fire
point of the bath fluid.
• Observe the limited working temperature range when using plastic bath
tanks.
• Never operate the unit without bath fluid in the bath.
• Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the fluid.
• Prevent water from entering the hot bath oil.
• Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (e.g., by switching
the unit on for a short moment).
• Use suitable connecting tubing.
• Avoid sharp bends in the tubing, and maintain a sufficient distance from
surrounding walls.
• Make sure that the tubing is securely attached.
• Regularly check the tubing for material defects (e.g., for cracks).
• Never operate damaged or leaking units.
• Always turn off the unit and disconnect the mains cable from the power
source before performing any service or maintenance procedures, or before
moving the unit.
• Always turn off the unit and disconnect the mains cable from the power
source before cleaning the unit.
• Always empty the bath before moving the unit.
• Transport the unit with care.
• Sudden jolts or drops may cause damage in the interior of the unit.
• Observe all warning labels.
• Never remove warning labels.
• Never operate units with damaged mains power cables.
• Repairs are to be carried out only by qualified service personnel.
Some parts of the bath tank and the pump connections may become
extremely hot during continuous operation. Therefore, exercise particular
caution when touching these parts.
Caution:
The temperature controlling i.e. of fluids in a reactor constitutes normal
circulator practice.
We do not know which substances are contained within these vessels.
Many substances are:
• inflammable, easily ignited or explosive
• hazardous to health
• environmentally unsafe
i.e.: dangerous
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators