9.1.
MENU PUMP – Setting of pump pressure
Factory setting:
stage 1
The capacity of the circulating pump is set by adjusting the motor
speed
Settings: stage 1 ... 4
Flow rate:
11 ... 16 l/m
Pump pressure:
0,22 ... 0,45 bar
1. Press the
key. Menu >Pu< is indicated.
2. Press the
key. The set parameter flashes (example: > 2<)
3. Change the parameter by pressing
or
and confirm by
pressing the
key.
or
Press the
key if the parameter is to be retained.
The pump flow is pre-adjusted in the factory and can be
modified to suit user requirements.
• Using a screwdriver turn the screw (1) anti-clockwise by
360 °.
• Using flat pliers turn the marking of the slide (2) to the
desired position.
Tighten the screw.
Examples:
Internal applications in the bath
A 100 % internal bath circulation
(for large bath tanks)
B Reduced internal bath circulation
(for smooth surface of bath fluid)
External/internal applications
C 40 % external discharge,
60 % internal circulation
(for large bath tanks)
D 80 % external discharge,
20 % internal circulation
(for small bath tanks)
- FP50-MA Refrigerated and Heating Circulators FP40-MA Refrigerated and Heating Circulators FP35-MA Refrigerated and Heating Circulators F34-MA Refrigerated and Heating Circulators F32-MA Refrigerated and Heating Circulators F33-MA Refrigerated and Heating Circulators F25-MA Refrigerated and Heating Circulators F12-MA Refrigerated and Heating Circulators