beautypg.com

6 specifications – INFICON RSH-600 Rotary Sensor User Manual

Page 13

background image

1 - 5

IPN

1

5

3800

-G

RSH-600 Operating Manual

1.6 Specifications

Number of Crystals. . . . . . . . . . . . . . 6

Crystal Size . . . . . . . . . . . . . . . . . . . 0.550 in. diameter

Installation Aperture . . . . . . . . . . . . . 2.0 in. diameter

Overall Length

RSH-600 w/flat head . . . . . . . . . 8.19 in. (208 mm) plus in-vac length

RSH-600 w/angled head . . . . . . 8.62 in. (219 mm) plus in-vac length

Adjustable in Vacuum Length . . . 7.9 in. (200 mm) (standard)

13.8 in. (350 mm)

17.7 in. (450 mm)

21.3 in. (540 mm)

25.6 in (650 mm)

Power Requirement . . . . . . . . . . . . . 115 V (ac) @ 50 mA or 24 V (dc) @ 20 mA

Crystal Switching Method . . . . . . . . . Air Actuated @ 55 psi (4 kg/cm

2

) regulated

Cooling Method . . . . . . . . . . . . . . . . Water-cooled @ 5 L/m at 2 kg/cm

2

(28 psi)

Air and Water Connections. . . . . . . . (3) 1/4 in. quick connects

Operating Temperature . . . . . . . . . . 300°C max with water cooling and Standard

Head Cover
400°C max with water cooling and Copper
Head Cover

Weight . . . . . . . . . . . . . . . . . . . . . . . RSH-600 - Flat - 8.5 lb. (3.8 kg)