If you have a service-related question about the 2020ppbPRO, consult this manual
first. If you cannot find the answer in this documentation, contact INFICON

’s

Technical Support Department.

When you call, please have the following information ready:

1. A description of what happened and what you were doing when the

problem occurred.

2. Any corrective action that you have tried.

3. The exact wording of any messages that appeared on the display.

WARNING

Do not service the 2020ppbPRO in a hazardous location.

Troubleshooting

Situation

Cause

Action

Zero Air Error

Contamination of sample
lines, sample probe, inlet
filter or fittings before the
detector

Clean or replace contaminated
item

Span gas and zero air are
mixed up.

Ensure that clean air is used to
zero the 2020ppbPRO. If you
are using gas bags, mark the
zero air and span gas bags
clearly.

Ambient air is
contaminated.

If you are not sure of the
cleanliness of the ambient air,
use a commercial zero grade
air to zero the 2020ppbPRO.

Span Gas
Error

Span gas and zero air are
mixed up.

Ensure that clean air is used to
zero the 2020ppbPRO. If you
are using gas bags, mark the
zero air and span gas bags
clearly.

Span gas concentration is
too low.

Ensure that cal gas
concentration is accurate.