Table 2. averaging intervals and period stel, twa, And peak operation, Erasing or downloading data – INFICON 2020ppbPRO Portable Photoionization Detector User Manual

Page 56

background image

50

ISP-074-576-P1D

50

Figure 12. Interval time adjustment

Interval simultaneously displays the STEL, TWA, and PEAK reading along with the
instantaneous reading. The 2020ppbPRO

’s datalogger can store up to 200 hours of

1 minute sample entries. The interval you select will determine the period of
operation. At the end of the period, the datalogger will be full and you will see the
“Data Log Full” status.

Table 2. Averaging Intervals and Period

STEL, TWA, and PEAK Operation

Averaging

Interval

(example)

Hours of Operation to

Fill the Datalogger

15 s

50

1 m

200

2 m

400

5 m

1000

15 m

3000

The 2020ppbPRO

’s display can show four values in Interval mode: STEL, TWA, and

PEAK, plus the instantaneous reading.

Erasing or Downloading Data

There are two options for data manipulation under the “Erase/Download” Data
selection.

“Clear Data Log” deletes all readings in the 2020ppbPRO memory.

NOTE: Deleted information cannot be recovered. You should play back and print or

download the contents of the datalogger before deleting any information.