EN- 65

Sensistor ISH2000 Technical Reference Manual (TRM)

C 500

C 500 Factory configuration of the leak
detector parameters

Default parameters

Range and default settings of all Sensistor ISH2000 parameters.

Configuration

Parameter

Range

Factory default

Customer modification

Analysis Unit

Several Choices

PPM

APC Time A

0.0 - 6000.0 s

10.0

APC Time B

0.0 - 6000.0 s

0

APC Time C

0.0 - 6000.0 s

0

APC Time D

0.0 - 6000.0 s

0

Audio Base Frequency

Several Choices

400 Hz

Auto Range

ON/OFF

ON

Audio Ready Pulse

ON/OFF

ON

Audio Threshold
(Detection)

0 - 100 %

0 %

Audio Threshold
(Analysis)

0 - 100 %

4 %

Brightness

0 - 21

21

Calibration Time

Min Calibra. Time-30 s

10 s

Clock

hh:mm:ss

-

Contrast

0-20

10

Correlation Values

1.00E-37 - 1.00E+37

1.00E+00 = 1

Date

YY-MM-DD

-

Debug Mode

ON/OFF

OFF

Detector Signal Level

0-100 %

20 %

Direct sensitivity
adjustment

ON/OFF

ON

Display Threshold

0-100 %

4 %

Invert Colors

ON/OFF

OFF

Language

Several choices

English

Measure/Print Button

ON/OFF

OFF

Min Calibration Time

0 - 30 s

5 s

Min Presentation Time

120 s

1 s

Multipoint Analysis

Several Choices

OFF

Multipoint Analysis Time

0.0 - 30.0 s

5.0 s