Gas sensing specification (in analysis mode), A 900 – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual
Page 25
EN- 25
Sensistor ISH2000 Technical Reference Manual (TRM)
A 900
Gas sensing specification (In Analysis Mode)
Selected unit
Sensitivity
Measurement
range
Linearity
Repeatability
ppm (H
2
)
0.5 ppm
0.5 - 2000 ppm
(0.2%)
Typ. ± 15% of reading
(within 0.1 - 10 x
calibration point in
range 0.5 - 100 ppm)
Typ. ± (10% of
reading + 0.3 ppm)
mbarl/s air
(using 5% H
2
/
95% N
2
as
Tracer Gas)
5 x 10
-7
mbarl/s
5 x 10
-7
- 4 x 10
-2
mbarl/s
Typ. ± 15% of reading
(within 0.1 - 10 x
calibration point in
range 1 x 10
-5
- 2 x
10
-
3
mbarl/s)
Typ. ± (10% of
reading + 3 x 10
-7
mbarl/s)
g/a R143a (using
5% H
2
/95% N
2
as Tracer Gas)
0.2 g/a
0.2 - 8300 g/a
Typ. ± 15% of reading
(within 0.1 - 10 x
calibration point in
range 0.2 - 420 g/a
Typ. ± (10% of
reading + 0.1 g/a)
See also other documents in the category INFICON Equipment:
- TGF10 Tracer Gas Filler (36 pages)
- Sensistor ILS500 F Leak Detection Filler (90 pages)
- T-Guard Leak Detection Sensor (85 pages)
- T-Guard Leak Detection Sensor Interface description (40 pages)
- Sensistor ISH2000P Hydrogen Leak Detector, Panel Model (51 pages)
- Sensistor ISH2000 HySpeed Hydrogen Leak Detector (54 pages)
- LDS3000 Modular Leak Detector (52 pages)
- LDS3000 Modular Leak Detector Interface description (56 pages)
- BM1000 Bus module (14 pages)
- I/O1000 I/O module (18 pages)
- CU1000 Control unit (24 pages)
- Helium Leak Detector Modul1000 (130 pages)
- Helium Leak Detector Modul1000 Interface description (40 pages)
- UL5000 Dry Helium Leak Detector (108 pages)
- UL5000 Dry Helium Leak Detector Interface description (14 pages)
- UL1000 Fab Dry Helium Leak Detector (119 pages)
- HLD6000 Refrigerant Leak Detector (76 pages)
- HLD6000 Refrigerant Leak Detector Interface Description (40 pages)
- IO1000 I/O module (18 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector (92 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector Interface description (36 pages)
- Sensistor XRS9012 Hydrogen Leak Detector User Manual (28 pages)
- Sensistor XRS9012 Hydrogen Leak Detector Maintenance manual (14 pages)
- Extrima Ex-certified Hydrogen Leak Detector (62 pages)
- Sensistor ILS500 Leak Detection System (107 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (58 pages)
- Sensistor Sentrac Hydrogen Leak Detector (86 pages)
- Protec P3000(XL) Helium Leak Detector (132 pages)
- Pilot Plus Vacuum Gauge (2 pages)
- CO Check Carbon Monoxide Meter (2 pages)
- GAS-Mate Combustible Gas Leak Detector (12 pages)
- Whisper Ultrasonic Leak Detector (8 pages)
- Vortex AC Refrigerant Recovery Machine 115V (20 pages)
- Vortex AC Refrigerant Recovery Machine 230V (16 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (2 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (44 pages)
- D-TEK CO2 Refrigerant Leak Detector (12 pages)
- TEK-Mate Refrigerant Leak Detector (12 pages)
- Compass Refrigerant Leak Detector (12 pages)
- D-TEK Select Refrigerant Leak Detector (12 pages)
- Explorer Portable Gas Chromatograph (369 pages)
- MicroFID II Portable Flame Ionization Detector (89 pages)
- DataFID Portable Flame Ionization Detector for Landfill Emissions Monitoring (91 pages)
- Hydrostik Hydrogen Fuel Cylinder Installation (7 pages)