A 800 sensistor ish2000 - technical specifications, Power supply specifications, Input and output connections – INFICON Sensistor ISH2000 Hydrogen Leak Detector User Manual
Page 23: A 800
EN- 23
Sensistor ISH2000 Technical Reference Manual (TRM)
A 800
A 800 Sensistor ISH2000 - Technical specifications
Power supply specifications
Input and output connections
Power
Sensistor ISH2000
Sensistor ISH2000C
Sensistor ISH2000P
AC mains voltage
100-240 V 50/60Hz.
-
AC mains voltage to the
charger
100-240 V50/60 Hz
-
AC mains current
Typically 1 A (2 A pulse at
power on).
-
AC mains current to the
charger
Typically 300 mA
-
Fuse
2 A slow/ 250 VAC.
-
-
Nominal battery voltage
-
16.1 VDC (nominal). The unit
will shutdown at approx.
12VDC, and will be opera-
tional at approx. 14VDC
-
Operating time
-
8 h
-
Charging time
-
6.5 h
-
Power supply voltage
-
-
24 VDC
Power supply current
-
-
3 A max
Type
Sensistor ISH2000
Sensistor ISH2000C
Sensistor ISH2000P
General
Power input connector
AC input connector, IES 320.
Charger input connector, 2.1
x 5.5 mm std. Positive centre.
4 pin Phoenix MC 1.5/5.81
Series Detachable screw ter-
minal.
Probe control port inputs
Minimum pulse length
40 ms
-
40 ms
Input impedance
50k ohm
-
50k ohm
Input maximum range
-34 to +38 VDC
-
-34 to +38 VDC
Input high
> 12.0 VDC
-
> 12.0 VDC
Input low
< 8.0 VDC
-
< 8.0 VDC
Probe control port outputs
Output current
max 0.5 A/output, max 2.5 A
total
-
max 0.5 A/output, max 2.5 A
total
Inductive loads
External clamp diodes
recommended
-
External clamp diodes rec-
ommended
Low stage voltage
Max 1.5 VDC
-
Max 1.5 VDC