3 RS-232 interface

15

In

te

rface D

e

scr

ipti

o

n

Ecotec E300

0,

k

ins

22e1

-j

, 1401

3

RS-232 interface

The Ecotec E3000 is equipped with a RS-232 interface which is located on the rear right
side of the instrument. This interface is of the DCE type (Data Communications
Equipment) and allows the connection of a PC for monitoring and data logging. The
connection is provided through a commercially available Sub-D plug.

The RS-232 protocol can be set to “ASCII”, “Diagnostics”, “Printer auto” or “Printer manual”.
The ASCII protocol is a protocol similar to the SCPI, a standard protocol widely used for
measuring equipment. The Diagnostics protocol is a binary protocol used for analysis in
quality control and trouble shooting during manufacturing and service.

3.1

Connecting the RS-232 interface

A standard RS-232 cable can be used (straight-through connecting cable, RxD and TxD not
crossed). The RS-232 hardware handshake must be switched off (in RS-232 control
program written by the user). If switching off of the hardware handshake is not possible,
an RS-232 connecting cable wired as follows may also be used:

Fig. 6: RS-232 cable

The levels on the RS-232 interface are defined as follows.

Table 8: RS-232 Pin assignment

Pin Name Signal

2

RXD

Receive data (Ecotec E3000

 PC)

3

TXD

Transmit data (PC

 Ecotec E3000)

5

GND

Reference ground

Table 9: RS-232 Level definition

Level Low

(L)

High

(H)

Voltage range

-3 V … - 25 V

3 V … 25 V

Logic state

logical 1

logical 0

Level designation

Mark

Space