Replacing the sinter filter of the sniffer handle – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual
Page 73

8 Maintenance
73
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
Ecot
ec E3
000,
k
in
a
22en
1
-r
, 1
407
7 Close the sniffer probe with your finger. With the water conservation tip, you have to
also cover the opening on the side. You should then be able to feel a vacuum. If not
then there is a leak somewhere and you have to check the screw connection. The
plugwasher may also have fallen out.
8 Confirm the work,
see “Confirm maintenance”, page 65
9 Calibrate the Ecotec E3000,
see “6.4.2 Calibration”, page 37
8.2.5
Replacing the sinter filter of the sniffer handle
The sinter filter is in the sniffer handle. To replace filter inserts, proceed as follows:
1 Turn off the Ecotec E3000.
2 Unscrew the two cross-head screws that hold the sniffer probe in place.
3 Remove sinter filter including o-ring.
Fig. 55: Sinter filter in the sniffer handle
4 Check the filter for visible soiling.
5 Insert a new sinter filter with o-ring in the base of the filter tip.
6 Tighten the sniffer probe again.
7 Turn on the Ecotec E3000.
8 Close the sniffer probe with your finger. You should be able to feel the vacuum. If not
then there is a leak somewhere and you have to check sniffer probe and handle.
9 Confirm the work,
see “Confirm maintenance”, page 65
10 Calibrate the Ecotec E3000,