3 replacing mains fuses, Replacing mains fuses – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual
Page 71

8 Maintenance
71
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
Ecot
ec E3
000,
k
in
a
22en
1
-r
, 1
407
8 Place the plastic container in the bore with the soaked liner first and close it with the
aluminum plug.
9 Tighten the plastic screw again. Ensure that the o-ring is placed correctly in the groove
of the plastic screw and that the bore is lined up properly.
10 Confirm the work,
see “Confirm maintenance”, page 65
8.2.3
Replacing mains fuses
The fuses are behind a flap near the mains plug. They are in two slots.
The mains fuses are available with order no. 200 000 914. Two identical fuses must be used
at all times.
1 Lever out the cover of the mains plug to the right with a screwdriver.
Fig. 51: Levering out cover to the right
2 Pull out the two slots and replace the fuses.
Fig. 52: Slot with fuse
3 Push the slots back in. Ensure that the arrows point up.
4 Close the flap.
Danger to life due to electric shock
Inside the device there are voltages of considerable magnitude. Touching parts where
electric voltage is applied can result in death.
►
Disconnect the device from the power supply prior to any maintenance work. Make
sure that the electric power supply cannot be reconnected unauthorized.