If the measuring value limit is exceeded, the display switches from green to red. You can
additionally set one speaker in the handle to output a signal and the LEDs in the handle to
flash or shine with maximum brightness,
see “6.3 Settings before measurements”, page 28
4.4
Technical data
Table 2: Technical data
Mechanical data
Dimensions (W
H D)
610 mm
370 mm 265 mm
Weight
34 kg
Ambient conditions
Permissible ambient temperature (during operation)
10 °C to 45 °C
Permissible storage temperature
-20 °C to 60 °C
Max. relative humidity up to 31 °C
80 %
Max. relative humidity from 31 °C to 40 °C
Linearly dropping from 80 % to 50 %
Max. relative humidity above 40 °C
50 %
Pollution degree
II
(according to IEC 61010/part 1: “Usually, only non-
conducting soiling may occur. However, temporary
conductivity caused by condensation is tolerable at times.“)
Max. altitude above sea level
2000 m
Electrical data
Supply voltages and frequencies
Country-specific, see rating plate on device rear
Power consumption
300 VA
Protection class
IP 20
Overvoltage category
II
Mains fuse
2
4 A slow-blow
Power supply cable
2.5 m
Noise level
< 54 dBA
Physical data
Minimum detectable leak rate
R134a
0.05 g/a (0.002 oz/yr)
R600a
0.05 g/a (0.002 oz/yr)
Helium
< 1
10
-6
mbar l/s
Measurement range
6 decades
Detectable masses
2 to 200 amu
Mass spectrometer
Quadrupol mass spectrometer
Ion source
2 cathodes
Time constant of the leak rate signal
< 1 s
Gas flow through the capillary
Measured at 1 atm (1013 mbar) at sea level. The flow rate
changes with geographical height and barometric pressure.
120 to 200 sccm
Time until ready for operation
< 2 min