7 Warnings and error messages
57
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
Ecot
ec E3
000,
k
in
a
22en
1
-r
, 1
407
W35 Flow through capillary is too low
(The lower limit value is fallen short of
longer than 2 s.)
The lower limit value is set too high.
Lower the lower flow rate limit value,
The capillary filter in the sniffer line is clogged.
Replace the capillary filter,
.
The sinter filter in the sniffer handle is clogged.
Replace the sinter filter,
.
The capillary is clogged.
– Replace the sniffer line.
Recalibrate the device,
.
– Replace the multi-function cable of the sniffer line. *
Recalibrate the device,
.
The internal filters in the main unit are clogged
Replace the internal filters (three pieces).*
Confirm the work in the service menu.
Recalibrate the device,
E36 Flow through capillary too high.
(The upper flow rate limit value is
exceeded longer than 2 s.)
The sniffer line is incorrectly connected.
Check the connection.
The upper limit value is set too low.
Set the upper limit value for the flow higher,
The capillary is broken or leaking.
– Replace the sniffer line.
Recalibrate the device,
.
– Replace the multi-function cable of the sniffer
line.Replace the multi-function cable of the sniffer
line.*
Recalibrate the device,
.
E37 Capillary broken! (>10 s)
(The upper limit value is exceeded longer
than 10 s. The emission is switched off to
protect the cathodes.)
The upper limit value is set too low.
Set the upper limit value for the flow higher,
The capillary is broken or leaking.
– Replace the sniffer line.
Recalibrate the device,
.
– Replace the multi-function cable of the sniffer
line.Replace the multi-function cable of the sniffer
line.*
Recalibrate the device,
.
E38 Capillary broken! (>60 s)
(The upper limit value is exceeded longer
than 60 s. The pumps are switched off to
protect the cathodes.)
The upper limit value is set too low.
Set the upper limit value for the flow higher,
The capillary is broken or leaking.
– Replace the sniffer line.
Recalibrate the device,
.
– Replace the multi-function cable of the sniffer
line.Replace the multi-function cable of the sniffer
line.*
Recalibrate the device,
.
E39 Emission failed
(The emission on both cathodes cannot
be switched on.)
– If the device was switched off for a longer period
of time, this error can occur during the first
10 minutes after start-up.
– Both cathodes are defective.
– The Transpector is defective.
Confirm the error message and restart the device. If the
problem still exists: Contact the INFICON customer
service.
Table 10: Warning and malfunction messages (cont.)
No. Message
Possible
error
sources
Fault clearance