INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector User Manual
Page 64

64
8 Maintenance
Tr
an
sl
ation
o
f the
o
rig
in
al
operatin
g
in
str
u
ct
io
ns
Ecot
ec E3
000,
k
in
a
22en
1
-r
, 1
407
Fig. 42: Calibration list
To have the complete calibration information displayed, select the appropriate line and
press the key with the magnifying glass. The calibration information includes:
Calibration mode (internal or external); for internal: gas in calibrated leak
Gas
– Internal calibration: one or more gas numbers
– External calibration: gas number, mass position, gas
Date and time of calibration
Number of operating hours at the time of calibration
Calibration factor
Peak position (if different from mass position)
Flow through sniffer line at the time of calibration
Cathode that was in use at the time of calibration (A/B)
Size of the calibrated leak that is used (external calibrated leak for external calibration
and ECO-Check for internal calibration)
Argon flow and mass deviation at the time of calibration
Flow of the gas that was calibrated and background signal
Maintenance intervals
To call up the elapsed operating hours of the device since the start-up, select
“Maintenance intervals”. The information does not apply to the sniffer line because
different lines could have been used.
The remaining operating hours up until the next maintenance work are displayed below.
The times below “Next maintenance for …” are based on confirmations that you enter
after maintenance work,
see “Confirm maintenance”, page 65
.