Installation
4.1
Unpacking
Unpack the T-Guard™ Leak Detection Sensor immediately after receipt even if it is
to be put into operation at some later date.
Examine the shipping container for any external damage.
Completely remove all packaging materials.
Notice
The three hose connectors are protected by plugs. Please only remove the
plugs when you connect the hoses to the T-Guard™.
Notice
Retain the shipping container and the packaging materials. Perhaps it will
be required in the event of claims for compensation.
Check if the T-Guard™ Leak Detection Sensor is complete (see Section
) and
carefully subject it to a visual inspection. If any damage is discovered, please
immediately inform the forwarding agent and the insurers. If it is necessary to
exchange the damaged part, please contact our orders department.
After unpacking T-Guard™ Leak Detection Sensor, you have to care for a proper
installation:
4.2
Installing Mechanically
Notice
T-Guard™ Leak Detection Sensor must be isolated from any vibration.
Vibration may lead to false measurement results!
Caution
Ensure sufficient ventilation.
In order to ensure adequate ventilation of the T-Guard™ Leak Detection Sensor, a
gap of at least 20 cm must be provided on top. Avoid heat sources being in the
vicinity of the T-Guard™ Leak Detection Sensor.
Caution
Do not block the ventilation inlets and outlets of the T-Guard™. Doing so could
damage the device.
Caution
Do not forget to attach the filters on the ends of the test and reference lines. They
protect the hoses from clogging and the sensor from dust contamination.