beautypg.com

1 cleaning and inspecting the pump heads – Heidolph ROTAVAC Vario pumping unit User Manual

Page 51

background image

page 51 of 61

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011

Use open-ended wrench (w/f 17) to loosen the union nut at

the hose connection next to the gas ballast.

11.1 Cleaning and inspecting the pump heads

Use open-ended wrench (w/f 17) to loosen the union nut at

the hose connection next to the inlet and remove hose.

Use Phillips screw driver to remove both screws at the fix-

ing plate of the controller and remove controller.

Remove control cable at the rear side of the controller. Do

not cant when removing plug connection!

Tools required (metric):

- diaphragm key w/f 46 (enclosed in the set of seals)

- open ended wrench w/f 14 / 17

- hex key size 4

- Phillips screw driver size 2

Please read section ”Replacing diaphragms and valves” completely before

starting maintenance.

Partially the pictures show pumps in other versions. This doesn’t influence replacing diaphragms and

valves of the pump.