Fault

Possible cause

Remedy

❑ Internal venting valve

does not respond, valve

symbol is not displayed.

External pressure transducer

connected and active?

✔ Select internal pressure

transducer or connect exter-

nal venting valve.

❑ Venting valve does not

respond, valve symbol is

displayed.

Valve soiled?

✔ Clean valve.

❑ Function ”Vac control”:

Control stops, ”arrow up”

is flashing.

Preset maximum pressure

exceeded?

✔ Confirm by pressing key

START/STOP. Adapt

maximum pressure value if

necessary.

❑ Function ”Pump down”:

Control stops, ”arrow

down” is flashing.

Pressure below preset mini-

mum pressure?

✔ Confirm by pressing key

START/STOP. Adapt

minimum pressure value if

necessary.

❑ Function is not displayed,

no menu.

No controllable device con-

nected (valve, pump)?

✔ Connect devices or use con-

troller as measuring device.

❑ Controller does not

respond when pressing

keys (except ON/OFF).

PC symbol is displayed.

Controller in remote mode?

✔ Control controller via inter-

face or switch off remote

mode.

❑ Controller does not

respond when operat-

ing any keys. No change

after switching off/on.

✔ Contact local distributor.

❑ Pump does not start or

stops immediately. Warn-

ing triangle and pump

symbol are flashing.

Pump has been exposed to

condensate?

✔ Allow pump to run for some

minutes at maximum speed

with atmospheric pressure

at the inlet.

Overpressure in the exhaust

line?

✔ Open exhaust line, ensure

a free (pressureless) outlet

line.

❑ Pump too noisy.

Loud exhaust noise?

✔ Connect hose or silencer to

pump outlet.

Diaphragm crack or dia-

phragm clamping disc loose?

✔ Perform maintenance.

Other than above mentioned

causes?

✔ Contact local distributor.

❑ Pump seized.

✔ Contact local distributor.