Example for use – Heidolph ROTAVAC Vario pumping unit User Manual

Page 33

background image

page 33 of 61

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011

The last process is stored in the temporary data memory as long as the controller stays switched on.

This program can be transferred to a storage space and edited.

Once the program is finished, the clock symbol starts to flash. Confirm the end of the program by press-

ing START/STOP (clock symbol will disappear).

Attention: If ”Autostart” is set to On, the program will start again (time will be reset to 00:00:00)

after a power failure or after switching the controller off/on. Only if the end of the program (clock symbol

flashing) has been confirmed by pressing START/STOP, the program will not start again.

Attention: If the controller is set to ”Defaults”: On”, all stored programs will be deleted.

Example for use

Example: Degassing and automatic distillation with timing at a rotary evaporator:

Program step 1 should be always a definite initial state, here atmospheric pressure (ATM). To reach this

state definitely, set a tickmark at ”Vent.” and ”Step” (press selection knob).

In step 2 pumping down starts as fast as possible (”Step”) to 300 mbar and the vacuum is kept for 10

minutes (degassing of the solvent).

In step 3 ”Auto ” causes an automatic search of the boiling vacuum in the pressure interval between

300 mbar and 2 mbar and automatic adaptation to a changing boiling pressure. The following step starts

if time is over (after 60 minutes) even if the preset pressure (2 mbar) is still not reached or if a vacuum

of 2 mbar is reached even if the preset time has not passed.

Step 4 vents to atmospheric pressure as fast as possible and switches off the control after one minute.

Program

No hh:mm:ss Vac Vent. Step Auto

01 00:00:00 ATM ✔ ✔

02 00:10:00 300

03 01:00:00 2

*

04 00:01:00 ATM ✔ ✔

05 00:00:00 0

06 00:00:00 0

07 00:00:00 0

08 00:00:00 0

09 00:00:00 0

10 00:00:00 0

- - - - - - - - - - - - - back - - - - - - - - - - - - - -

01:11:00

* If the pressure difference between the vacuum for

degassing and the expected vacuum for distillation

is very small (distillation vacuum >75% of the de-

gassing vacuum), select the function ”Auto ” in-

stead of ”Auto ” (adapting the vacuum).