Fault

Possible cause

Remedy

❑ No display.

➨ Mains not plugged in?

✔ Plug in mains plug.

➨ Controller or pumping unit

switched off?

✔ Switch device on.

➨ BUS cable to internal power

supply not plugged in?

✔ Plug in BUS cable at control-

ler.

➨ Other causes (device defec-

tive)?

✔ Contact local distributor.

❑ Display disappears.

➨ Short circuit at connected

valves?

✔ Replace valves.

➨ Short circuit at the RS 232

plug?

✔ Check plug and cable.

➨ Other causes (device defec-

tive)?

✔ Contact local distributor.

❑ Pressure reading incor-

rect.

➨ Pressure transducer not

adjusted?

✔ Readjust the controller.

➨ Humidity inside the pressure

transducer?

✔ Let the pressure transducer

dry, e.g. by pumping. Read-

just if necessary. Determine

and eliminate the cause for

humidity.

➨ Pressure transducer contami-

nated?

✔ See „Cleaning the pressure

transducer“.

❑ Digital pressure reading

is flashing, display shows

”0.0”.

➨ Pressure transducer not

correctly adjusted under

vacuum?

✔ Adjust controller correctly.

❑ No digital pressure read-

ing.

➨ Pressure transducer defec-

tive?

✔ Contact local distributor.

❑ Digital pressure reading

is flashing, one blip*.

➨ Overpressure at the pressure

transducer, pressure > 1060

mbar?

✔ Release pressure immedi-

ately (risk of bursting)!

❑ Warning triangle and

coolant valve symbol are

flashing, four blips*.

➨ Coolant valve removed or

defective?

✔ Check connection cable of

the valve or use new valve or

reconfigure without valve.

❑ No digital pressure read-

ing, warning triangle is

flashing, seven blips*.

➨ Seven blips: Internal sensor

defective?

✔ Contact local distributor.

❑ Warning triangle and

pump symbol are flash-

ing, six blips*.

➨ Fault at the pump?

✔ Check pump, restart control-

ler.

❑ Clock symbol is flashing.

Control is stopped.

➨ Preselected process time is

over?

✔ Confirm by pressing key

START/STOP.

9 Troubleshooting

* Blips only if in function ”Configuration” sound is set to ”On”.