0 control off

1 pump down/

determining boiling point

2 set vacuum reached/

boiling pressure found

3 current pressure below set vacuum/

automatic switch-off

0 VACUULAN

1 Pump down

2 Vac control

3 Auto mode

4 Program

5 measurement device

0: venting valve closed

1: venting valve open

0: coolant valve closed

1: coolant valve open

0: in-line valve closed

1: in-line valve open

0: pump off

1: pump on

Description

Response

XXXXXX

XXXXXXXXX

Operation

Status

process control

Fault status

Command

IN_STAT

IN_ERR

Read commands

0: last interface command correct

1: last interface command incorrect

0: catchpot not full

1: catchpot full

0: no external fault

1: external fault

0: no fault at pressure transducer

1: fault at pressure transducer

0: no overpressure

1: overpressure

0: no fault at venting valve

1: fault at venting valve

0: no fault at coolant valve

1: fault at coolant valve

0: no fault at in-line valve

1: fault at in-line valve

0: no fault at pump

1: fault at pump