Environmental Express SPE-Express with Cover User Manual
Page 43
Environmental Express, Inc.
800.343.5319/843.881.6560
42
•
Click to open the “Hexane Bottle Rinse Sol”.
•
Once the n-Hexane is evacuated from the line (Nitrogen will be heard entering the n-Hexane
source bottle), close the Hexane solenoid by pressing “Hexane Bottle Rinse Sol”.
•
Click to open the “Methanol Rinse Sol”
•
Once the Methanol is evacuated from the line (Nitrogen will be heard entering the Methanol
source bottle), close the Methanol solenoid by pressing “Methanol Rinse Sol”.
•
Click to open the “DI Water Rinse Sol”
•
Once the DI Water is evacuated from the line (Nitrogen will be heard entering the DI Water source
bottle), close the DI Water solenoid by pressing “DI Water Rinse Sol”.
•
Repeat this process in the Manual Screen for all 3 stations to ensure all chemicals have been
evacuated from the tubing lines.
•
Carefully remove the tubes within the bottle caps from the sources bottles.
Caution: the tubes may contain some residual chemical.
•
Remove the top 3 fittings from the back of the manifold using the Fastener Tool supplied with the
unit.
•
Remove the bottom 3 tubes by twisting the blue portion of the luer fitting clockwise.
•
The bottle harness should be stored in a dry environment which has low risk of contact with
contamination.