Environmental Express SPE-Express with Cover User Manual
Page 20
SPE-‐Express Operation and Instruction Manual
www.envexp.com
19
2.3.6
Method Creation Screen
Press the “Method Creation” button on the Environmental Express SPE-Express Main screen or press F5.
The Method Creation Screen allows the user to create specific method parameters for testing.
2.3.6.1 Pre-Rinse
In this area, the user may enter parameters for hexane pre-rinse.
Use the arrow controls to:
•
Enter the desired hexane pre-rinse volume in milliliters.
•
Enter the desired hexane pre-rinse soak time in seconds.
•
Enter the desired hexane siphon time in seconds.
•
Enter the desired number of hexane pre-rinses.
2.3.6.2 Disk Activation
In this area, the user may enter parameters for disk activation.
Use the arrow controls to:
•
Enter the desired methanol injection volume in milliliters.
•
Enter the desired methanol soak time in seconds.
2.3.6.3 Sample Filtration
In this area, the user may enter the filter drying time for sample filtration.
2.3.6.4 Elution
In this area, the user may enter parameters for the elution step. Use the arrow controls to:
•
Select the desired hexane spray volume in milliliters per spray.