beautypg.com

Environmental Express SPE-Express with Cover User Manual

Page 31

background image

Environmental  Express,  Inc.

 

 

         800.343.5319/843.881.6560

 

 

30  

Press “Exit” to return to the Operator Test screen.

3.2.3

Draining a Station Manually

If a test is aborted prior to or during the bottle drain portion of the test the “Drain Station” button may be
pressed which will allow the sample bottle to drain into the waste container.
Caution: if the test aborted during the bottle drain procedure, sample water may be caught between the filter
pan and the ball valve. If this water is not removed by initiating the drain function it could leak while
unlatching the swinging assembly.

Press “Drain Station”.

A prompt will appear to confirm the initiation of the draining procedure.

Press “Ready”. The pathway to the waste container and ball valve directly below the bottle will open
allowing the sample water to drain.

Press “Exit” to return to the Operator Test Screen without initiating the Drain procedures.

Once “Ready” is pressed prompt will be removed and the “Draining Station” button on the Operator
Test screen will display red to indicate that this feature is on.

Press “Draining Station” to conclude the draining procedures.

3.2.4

Cleaning a Station Manually

If the hardware within the swinging assembly becomes clogged during a test the Manual Clean may be
used to clean the system without having to exit the Operator Test Screen. This allows a clogged station
to be cleaned without interrupting tests on other stations.

Before using a manual clean it is recommended to perform a Station Drain until all residual water has
entered the waste container.

Swing the assembly into the lowered position and leave the filter in place.

Press the Manual Clean button.