o Ensure doors remain closed
o Install a new filter disk
o Install a sample bottle
o Ensure bottle assembly is latched in the upright position.
•
Press “Ready” for the second test to begin or press “Exit” to return to the Operator Test screen if
subsequent samples are not scheduled.
•
Once the previous test completes the software will alert you with a prompt. The second test will not
proceed past the filter drying step until the previous test is completed. Press the radio buttons to
confirm that the following steps have been completed:
o Remove the aluminum evaporating pan from the heater block
o Place a new pre-weighed aluminum evaporating pan on the heater block
o Remove spring clamp then lower sliding glass chamber into the down position
o Remove and discard the drying cartridge
o Install a new drying cartridge
•
Press “Next” to continue to the next prompt. Press the radio buttons to confirm that the following
steps have been completed: