beautypg.com

Spe-­‐express operation and instruction manual – Environmental Express SPE-Express with Cover User Manual

Page 22

background image

SPE-­‐Express  Operation  and  Instruction  Manual  

 

 

www.envexp.com

         

 

 

21  

2.3.6.6 Method Parameter Specifications

Status

Name

Input Limits

Description

Pre-Rinse

Hexane Pre-Rinse Volume

10-20 mL

Cleans system below sample bottle with specified amount of

hexane, allows hexane to soak and then siphons hexane to hexane

waste container.

Hexane Pre-Rinse Soak Time

30-120 seconds

Hexane Siphon Time

60-120 seconds

Number of Hexane Pre-Rinses

1-5 rinses

Disk Activation

Methanol Injection Volume

10-20 mL

Injects desired amount of methanol onto the filter and allows

methanol to soak into the filter.

Methanol Soak Time

60-120 seconds

Sample Filtration

Filter Dry Time

10-20 minutes

After the sample bottle has drained, the time the filter is allowed to

dry with vacuum prior to elution.

Hexane Rinse Parameters

Spray Volume

3.1-30 mL/spray

Volume of hexane used to spray the sample bottle.

Sprays per Rinse Cycle

1-10 sprays

Number of sprays per rinse cycle

Total Spray Volume

9-30 mL

Total volume of hexane used to spray the sample bottle during 1

rinse cycle. *Automatically updates to show the total hexane usage

during the bottle rinses based on the Total Spray Volume and the

Number of Rinse Cycles.

Drain Time

30-120 seconds

Amount of time between each rinse cycle.

Number of Rinse Cycles

1-5 rinses

The number of rinse cycles the system completes during elution.

Total Volume

27-120 mL

The total volume of hexane used during the elution steps.

*Automatically updates to show the total hexane usage during the

bottle rinses based on the Total Spray Volume and the Number of

Rinse Cycles.

Vacuum Parameters

Vacuum ON Time

10-20 seconds

Pulsing ON/OFF of the vacuum within the glass chamber to remove

eluted hexane from the filter.

Vacuum OFF Time

5-20 seconds

Total Vacuum Cycles

5-10 cycles

Vacuum Dry Time

60-180 seconds

Final vacuum duration to pull all remaining hexane into the glass

chamber.

Nitrogen Purge

3-120 seconds

Application of nitrogen to push all hexane from the glass chamber

to the evaporation pan.

*automatically calculated

*automatically calculated