Environmental Express SPE-Express with Cover User Manual
Page 36
SPE-‐Express Operation and Instruction Manual
www.envexp.com
35
•
Methanol Rinse Solenoid
•
Deionized Water Rinse Solenoid
•
Hexane Chamber Rinse Solenoid
•
Vacuum Control Solenoid
•
Nitrogen Pressure Solenoid
The user may control how long one of the above solenoids is powered. Toggle the Timed Control Switch to
Auto.
•
Enter the desired time that the solenoid will remain active.
•
Press the button corresponding with the solenoid to be activated.
•
The Elapsed Time will count up to time completion and then deactivate the solenoid.
The user may power any one of the 3-Way Solenoid Valve Interface solenoids (the “Hexane Waste Sol”,
“Water Miscible Waste Sol” or “Eluted Sample Path Sol” by pressing the button corresponding with the
desired solenoid (cannot be completed automatically).
The user may power the sample release solenoid by pressing the corresponding button (cannot be
completed automatically).
3.6.2
Ball Valve
In this area the user can control the ball valve. Press the switch to open or close the ball valve.
3.6.3
Status Indicators
In this area, the user may see the current status of the door and latch. Press the switch to power the latch
and door switch.
3.6.4
Main Fans
In this area, the user may see the current status of the vent and exhaust fan. Press the “Main Fans” switch
to power the vent and exhaust fans
3.6.5
Sensor Chart
•
Press the Power Enable button at the lower left hand corner of the graph to turn on the
station sensor.
•
The graph will update the sensor status with respect to time. The graph will display one if
the sensor is in the ON state (the sensor will have an orange light near the cable). This ON
state should be seen when sample water is draining to the waste container. The graph will
display zero if the sensor is in the OFF state.
The ON/OFF states require occasional calibration to maintain accuracy. See Sensor Calibration
instructions, Section 5.5.
•
Press the power button again to turn off the station sensor.