DIGIWEIGH DWP-98 User Manual
Page 19
b. The quantity added to platter is less than 100% of the maximum
counts previously attained from the same transaction.
If the above requirements are fulfilled, a new unit piece weight will be
determined and displayed on the Unit Piece Weight Panel and confirmed
by 2 audio beeps.
Auto Unit Piece Weight Enhancement Function will terminate when: -
a. a zero weight is detected or during the transaction process, or
b. quantity added is more than 100% of the maximum counts previously
attained from the same transaction, or
c. quantity removed is not more than 50% of the maximum counts
previously attained from the same transaction.
To start Auto Piece Weight Enhancement Function again after
terminated, repeat 6.5.2.
6.7 MEMORY ACCUMULATION FUNCTION
6.7.1 To Accumulate a Transaction to Memory
6
7
8
a. When quantity is being displayed on Quantity Panel, press [M+]
9
to
and accumulate data of current transaction to memory,
b. The instrument: -
displays “ACC” on Weight Panel. ACC denotes data has been
accumulated to memory,
displays “n” on Unit Piece Weight Panel. n donates the number
of transactions which have been accumulated to memory.
displays Total Indicator and total accumulated value on
Quantity Panel.
6
Transaction with zero quantity will not be accumulated.
7
Transaction with weight less than 20d will not be accumulated.
8
All data stored will be erased when the instrument is powered off or changing
weight unit.
9
Unstable weight will not be accumulated to memory. If M+ is pressed when weight
is unstable, the instrument will reject this command and response with 2 beeps.