A Despatch Protocol Plus controller/high-limit device is used to control the unit. This is
located on the control panel of the oven.

The Protocol Plus controller provides temperature control for the process.

The Protocol Plus controller provides outputs for the cooling fan, door lock

switch/door release pushbutton, and optional beacon light.

As many as eight (8) profiles for oven heating cycles are stored in the Protocol

Plus controller. These are accessed by the operator using the Protocol Plus
keypad.

The Protocol Plus controls the solenoid valves in an inert atmosphere oven for

purge and maintain operation.

Optional MODBUS RS422/485 serial communications hardware may be installed

on the Protocol Plus controller, with a 9-pin communications port located on rear
of oven. This provides the ability to network the oven(s) with a host PC.

The optional three-color beacon light provides visual indication of the cycle

status, as follows:

Green – Cycle in process
Yellow – Standby. Operator has unloaded product. Oven waiting for next

cycle.

Red – Error or fault condition.

The optional High-Limit Alarm indicates when the high-limit setpoint has been

exceeded. An illuminated Alarm Silence switch lets the operator turn off the
alarm while providing a red warning light until the high-limit condition is corrected.
(See High-Limit description later in this manual.)