31
CALIBRATION CHECK
(See Configuration Mode, Appendix A, and Appendix B.)
1.
Power up the instrument and allow to stabilize for at least 5 minutes (RTD and
DC) or 30 minutes for T/C input.
2.
After the stabilization period, connect the appropriate input device and check a
number of input points.