Despatch MIC1162 Hi-limit User Manual
Page 26

22
SECTION 6: CONFIGURATION
 
All configurable parameters are provided in Table 6-1. This table illustrates the display 
sequence, parameter adjustment and factory setting for each step. 
 
Depression of the SCROLL key will cycle the display. 
 
To enter the Configuration mode, press and release the SCROLL key until ConF is 
displayed. Use the DOWN key to enter the Configuration mode. Depress and release 
the SCROLL key to sequence through the parameters and their values, alternately 
showing the parameter code in the lower display with the upper display blank, then the 
parameter code with the parameter value in the upper display. Use the UP and DOWN 
keys to adjust the parameter values. 
 
After adjusting a parameter, the upper display will flash, indicating that the new setting 
has yet to be confirmed. When the setting is as required, it may be confirmed by 
pressing the RESET key, and the upper display will stop flashing. After confirming a 
change, press the SCROLL key to proceed to the next parameter. 
 
The DOWN key may also be used to advance to the next parameter when a parameter 
code is showing in the lower display and the upper display is blank. This is a faster way 
of scrolling through the parameter list. 
 
To exit the Configuration mode, press the UP key whenever a parameter code is 
displayed in the lower display and the upper display is blank. The instrument will 
automatically exit the Configuration mode if no key is pressed in one minute. 
 
DEFAULT PARAMETER INDICATION 
 
If a parameter value, such as Input Select, was changed while in the Configuration 
mode, a decimal point after each digit will be lit when returning to the Control mode. 
This display indicates all Setup mode parameters have been set to their default 
condition. To clear this condition, enter the Setup mode and make a parameter value 
change and review each parameter for its proper setting. 
 