Apply power to the instrument and within 30 seconds of power-up, press and
hold the DOWN and SCROLL keys simultaneously for about 5 seconds. The
upper display will show ip_1 and the lower display will show CAL.


5.

Use the UP/DOWN keys as required to change the input type number as
required:


CAL INPUT NO.


INPUT TYPE

1


DC 0 - 50 mv

2


DC 0 - 10 V

3


DC 0 - 20 mA

4


RTD

5


Thermocouple


NOTE: If required, only one input type may be calibrated. EXCEPTION: If it is required
to calibrate the thermocouple input (Input Type 5), it is necessary to calibrate the DC 0 -
50 mV (Input Type 1) first.

6.

Press the RESET key, the upper display will show _ _ _ _.


7.

After a few seconds, the upper display will show ip_X, where X is the CAL INPUT
No., if the calibration was successful. If the upper display shows FAIL, the
calibration was not successful - check the jumper positions, wiring, CAL INPUT
No. and try again.


8.

To calibrate all inputs, repeat steps 2 through 7 for each of the other input types.


9.

The calibration procedure is now complete. Disconnect power, remove input
connections.