C.B.S. Scientific HTLE-7002 User Manual
Page 6

C. B. S. Scientific 6
HTLE-7002
*1.2 Specifications
Constructions:
Electrophoresis
base
Acrylic
Enclosure/lid
PVC/Acrylic
Buffer Tanks
Acrylic
Cooling block
Aluminum/stainless
steel
Air bag
Nylon/Delrin
Mains input
10A/250VAC/50Hz
or
20A/125V/60Hz
Fuse
5A/250V
(TT),
IEC
127
Power supply output
6A/250VAC
DC Power Input
20A/1000V
Power
cord
Continental Europe, CEE7/7
250VAC/6A, 65º C, 3 x 0.75mm
3
conductor UK & Ireland BS1363,
240VAC 50Hz, 5A, 70º C, 3x0.75mm
3
conductor U.S./PR, 16awg, 3 conductor,
60º C, SJT
Safety Certification
EN61010-1-1993
(IEC1010-1)
EN 50082-1 (IEC 801-2,3,4)
EN
55022
(CISPR
22)
1.3 Overall Sizes and Shipping weight
HTLE-7002
Shipping weight
69 lbs
Overall size
(inches) (l x w x h)
24 x 19 x 8
Overall size
(cm) (l x w x h)
61 x 48 x 20
1.4 Safety
Power to the HTLE-7002 is supplied by an external DC voltage power supply that must be
ground isolated so that the DC voltage output floats with respect to ground. For any power
supply used, the maximum specified operating parameters for
the units are:
Maximum Limits
100VDC/cm of gel
20W
watts
power
50mA
current
65
°C ambient temperature
Current to the unit, provided from the external mains connection, must enter the unit through
the safety cover assembly, providing a safety interlock to the user. Current to the power supply is
provided from the safety interlock. Both AC and DC is interrupted should the safety interlock be
activated. Current to the unit is broken when lid is lifted. Do not attempt to use the unit without
the safety lid, and always turn the power supply off before opening the lid, or when
working with the unit in any way. Follow safety precautions specified by the power
supply manufacturer.