Eppendorf Innova 40R Shaker User Manual
Page 46

Troubleshooting
New Brunswick Innova 40/40R Shaker
English (EN)
46
Excessive operating noise.
• Load out of balance.
Unload all contents and
reload.
• Loose components in platform,
subplatform, and/or drive
assembly.
Call for service.
Shaker does not reach set
temperature.
• Shaker is running in Program
mode (look for Run icon in
display).
• Shaker speed has been
changed by RS-232 command/
computer interface.
• Heater fuse blown.
• Compressor fuse blown.
Replace fuse.
• Compressor over-pressure
switch activated.
• Defective heater.
• Defective refrigeration system.
Call for service.
• Ambient temperature too high
or too low.
Adjust the room temperature.
Incorrect temperature indication.
• Temperature Offset has been
programmed.
Look for Offset icon in display.
• Defective RTD assembly.
• Defective main board.
Call for service.
Symptom
Cause
Solution