5 report request commands, Report request commands – Eppendorf Innova 40R Shaker User Manual
Page 42

Remote programming
New Brunswick Innova 40/40R Shaker
English (EN)
42
6.5
Report request commands
PW
Write Program step to
memory
PW<Space>
•
Arg1 = Program number
(1 – 4)
•
Arg2 = Step number (1 –
15)
•
Arg3 = Temperature
setpoint °C
•
Arg4 = Speed setpoint
rpm
•
Arg5 = CO
2
setpoint %
•
Arg6 = Step
(0 – 99)
•
Arg7 = Step
min (0 – 59)
Code Meaning
Command
format
Return data
RI
Report software version
RI<CR>
•
Arg1 = Display module software
version
•
Arg2 = Control module software
version
RP
Report parameter list (ASCII text header)
RP<CR>
•
Arg1 = Speed rpm
•
Arg2 = Temperature °C
•
Arg3 = CO
2
%
RV
Report measured values
RV<CR>
•
Arg1 = Speed rpm
•
Arg2 = Temperature °C
•
Arg3 = CO
2
%
RS
Report setpoint values
RS<CR>
Code
Meaning
Command format
Return data
This manual is related to the following products: